Variable Reluctance Laboratory Pressure Transducer Description:

The DP15 Variable Reluctance Laboratory Pressure Transducer is a general purpose differential pressure transducer that features field replaceable sensing diaphragms so that the full scale may be changed anywhere between +/-0.08 psid and +/-3200 psid. The DP15 Variable Reluctance Laboratory Pressure Transducer is the most cost-effective way to make pressure measurements in the laboratory where pressure measurement requirements are constantly changing: the DP15 Variable Reluctance Laboratory Pressure Transducer has 23 different diaphragm ranges available. When used with a Validyne carrier demodulator, the DP15 Variable Reluctance Laboratory Pressure Transducer can produce a +/-10 Vdc analog output proportional to the full scale pressure range. The rugged construction of the DP15 makes it highly resistant to shock and vibration. Because the diaphragm is the only moving part, the frequency response of the DP15 Variable Reluctance Laboratory Pressure Transducer to pressure pulsation's is exceptionally high.

The DP15 Variable Reluctance Laboratory Pressure Transducer will accept liquids and gases on both sides, and is constructed of 410 stainless steel. In addition, nickel and gold platings is available for aggressively oxidizing media. The O-ring seals are available in a variety of compounds. Bleed screws in each cavity facilitate liquid filling for dynamic pressure measurements.

Variable Reluctance Laboratory Pressure Transducer Applications:

  • Laboratory Pressure Measurement

  • Air & Liquid Flow

  • Fluid Level

  • Pressure Transient Studies

Variable Reluctance Laboratory Pressure Transducer Specifications:

Ranges: +/-0.08 to +/-3200 psid
Accuracy: +/-0.25% FS include non-linearity, hysteresis, and non-repeatability
Overpressure: 200% FS , to 4Kpsi
Line Pressure: 3200 psig
Output: +/-35 mV/V, nominal
Inductance: 20 mH, each coil
Zero Balance: Within +/-5 mV/V
Excitation: 5Vrms, 3KHz to 5khz
Pressure Media: Liquids and Gases compatible with 410 SST
Temperature Range: 0ºF to 165ºF, standard, -65ºF to +250ºF, extended
Temp Effect on Zero: 1% FS/100ºF, typical
Temp Effect on Span: 2%/100ºF, typical
Pressure Connection: 1/8 inch female NPT
Electrical Connection: PT02A-10-6P Bendix or equal

Diaphragm Range Code Chart - DP15

Available Ranges -20 to -64

Range Code Psi In Hg In H2O KPa Torr CM H2O
20 .125 .25 3.5 .86 6.5 8.80
22 .20 .41 5.5 1.40 10.3 14.0
24 .32 .65 8.9 2.2 16.5 22.5
26 .50 1.02 14.0 3.5 25.8 35.0
28 .80 1.6 22.2 5.5 41.4 56.0
30 1.25 2.5 35.0 8.6 65.0 88.0
32 2.0 4.1 55.0 14.0 103 140
34 3.2 6.5 89 22.0 165 225
36 5.0 10.2 140 35.0 258 350
38 8.0 16.0 222 55.0 414 560
40 12.5 25.0 350 86.0 650 880
42 20 41.0 550 140 1030 1400
44 32 65.0 890 220 1650 2250
46 50 102 1400 350 2580 3500
48 80 160 2220 550 4140 5600
50 125 250 3500 860 6500 8800
52 200 410 5500 1400 10300 14000
54 320 650 8900 2200 16500 22500
56 500 1020 14000 3500 28500 35000
58 800 1600 22200 5500 41400 56000
60 1250 2500 35000 8600 65000 88000
62 2000 4100 55000 14000 103000 140000
64 3200 6500 89000 22000 165000 225000